← back to 2026-10-31

Short version: three networked Rigol instruments, a power supply, an electronic load and an oscilloscope, driven as one by 462lxi, our command line tool. Together they can source, absorb and watch almost anything a small electronics project needs. A certified multimeter then showed that the instrument we trusted most was the one that was wrong, and that a power supply channel can sink current while switched off.

The instruments

Role Instrument Ratings
Power supply Rigol DP832A 3 isolated linear channels: 2 × 30 V / 3 A, 1 × 5 V / 3 A
Electronic load Rigol DL3031A 150 V, 60 A, 350 W; CC, CV, CR, CP
Oscilloscope Rigol MSO1074Z 4 analog channels, 70 MHz, 1 GSa/s, 8 bits
Reference RISH Multi 18S certified true-RMS multimeter, used to check the others

The first three speak LXI: SCPI commands over the network, through VXI-11. The multimeter is handheld, for now: its infrared link to the PC is on the to-do list.

One tool to drive them

462lxi is a single Python file with no dependency: no PyVISA, no vendor library. It implements VXI-11 directly over ONC RPC, which takes fewer than two hundred lines. It reads values the way an electronics engineer writes them (3v3, 4u7, 2k2, 500mA), and every command can answer in JSON for scripts.

462lxi devices                      # are all three online?
462lxi psu set 1 3v3 500mA --on     # voltage and current limit, then on
462lxi load set cc 1A --on          # load in constant current
462lxi scope wave 1 trace.csv       # 1200 points, acquisition keeps running
462lxi --json psu meas              # for scripts

Most of its code is not about sending commands, it is about not trusting the answers. Each instrument handles a bad value in its own silent way: the supply ignores an out-of-range setpoint and replies No error; the oscilloscope clamps it, shows a popup, and loses the next request; the load, to its credit, refuses with a proper error but resets its setpoint to zero on every range change, which in CR mode is a short circuit. So 462lxi checks every value against the limits it reads from the instrument, changes load ranges only with the input off and the setpoint at its minimum, and reads back what it wrote.

What the bench can do

Source. Three channels from 0 to 30 V (5 V on the third), up to 3 A each, with 1 mV / 1 mA setpoint resolution and 0.1 mV / 0.1 mA readback. Two channels in series give 60 V, or ±30 V around a common point. Over-voltage and over-current protections per channel. The supply cannot absorb current: it is a source only.

Absorb. Up to 150 V, 60 A and 350 W in constant current, voltage, resistance or power. The load also has a dynamic mode (current steps up to 30 kHz, slopes up to 5 A/µs), a 512-step list mode and a battery discharge mode with capacity, energy and duration counters. Its Sense terminals measure at the device under test, so the 50 mΩ of bench cables (50 mV at 1 A) drop out of the measurement.

Observe. Four analog channels at 70 MHz, plus logic channels and a signal generator on this model. 462lxi reads the displayed trace (1200 points in about 50 ms, without stopping acquisition) rather than screenshots, which is 20 to 40 times faster for repeated measurements.

Combine. This is where the bench earns its keep. Each important quantity can be read by two instruments, which checks the whole chain.

Test Instruments How
Efficiency and regulation of a converter supply + load input voltage and load swept point by point, both sides measured
Switch-mode supply validation DUT + load + scope load in dynamic mode, scope on the output: ripple and transient response
Power-on edge supply + scope scope armed in single shot, then the channel turned on
Current limit check supply voltage high, current capped: the channel must switch to CC
Real battery cell supply + load CC/CV charge, pulsed discharge, capacity and model (see the battery post)
Simulated battery supply (+ load) the supply follows a cell model, measured or generic, new or aged

A test is a script: it can run unattended for days, write every point to a text file, and be run again identically.

What the multimeter found

Two days ago, we calibrated the load in software, taking the supply as the reference. This week we checked both against the certified multimeter, in voltage and in current. The supply was the one off:

  • channel 2 reads its own voltage 10 to 12 mV high (within its specification of 0.05 % + 10 mV), and its current 1.2 mA + 0.09 % high, even at zero current;
  • its output sits 2 to 14 mV below the setpoint, and not linearly, so correcting the setpoint is not enough: the output has to be regulated on a measurement;
  • the load, raw, was already right within ±2 mV. Most of the “7 mV + 0.11 %” error we had corrected last time was the supply’s. Its +43 mA current offset in CV mode, 15 V range, on the other hand, is real: the multimeter sees it too.

462lxi now corrects both instruments from the multimeter readings. After correction, they agree with it within ±2 mV and ±0.4 mA. This is not a traceable calibration, but the reference is a certified instrument, not another bench instrument.

The channel that sinks 40 mA

The most surprising finding: a DP832A channel absorbs a constant 37 to 41 mA as soon as the voltage on its terminals is above its setpoint, output on or off, whatever the current limit. It does not show it in its own readback. Connect a charged cell to a channel that is switched off, and the cell slowly discharges into it. With a weak source (under 10 mA) it pulls the node down to about 0 V, which looks exactly like a wiring short.

The only inert state we found: channel on, setpoint above the node, current limit at 0 A. Even then it supplies about 1.2 mA. Until we have relays to isolate the outputs, the rule on the bench is simple: unplug a cell from any channel that is not working.

Two channels, one 60 V supply

The DP832A has no internal series mode: its first two channels are isolated from each other, and one wire from CH1− to CH2+ puts them in series. CH2 goes at the bottom, because CH2 and CH3 share their negative inside the instrument (the manual forbids putting those two in series or in parallel). The 5 V channel then stays referenced to the negative of the whole assembly.

462lxi drives the pair as a single channel, 1+2. Each channel gets half the setpoint and the same current limit, which the manual requires, and the measured voltages are added. The current is read from channel 2, the one corrected against the multimeter, since the same current flows through both.

462lxi psu batt 1+2 1Ah 200mA --cells 12 --on
462lxi psu charge 1+2 10Ah 2A --cells 13 --on

A guard checks the pair at every reading and stops on any of these:

  • one channel switching off alone;
  • a channel above its own setpoint, which means it is sinking;
  • the two current readings disagreeing, the other sign of a channel sinking;
  • current limiting, in modes where it is not expected.

On the bench, with the load as a test resistor:

  • a simulated 12S cell at 50.34 V: the two halves within 0.2 mV of each other, and the sum within 2 mV of the load’s sense and the multimeter;
  • the same in battery-simulation mode, with the load absorbing 300 mA at the simulated voltage: 13S, 50 V, 15 W, held to 1 mV for one minute;
  • load set above the current limit: both channels off within 1.5 s;
  • channel 1 switched off by hand during a run: channel 2 off at the next reading, 0.25 s later.

The first 13S run stopped for nothing. A channel sitting exactly on its setpoint reported itself “unregulated” while sinking nothing: the current readings from the supply and the load agreed. When current limiting is expected, the guard now relies on the voltages and the currents, not on that status.

Two findings came out of the session:

  • Two channels switched off in series both sink. About 38 mA flows through the pair, and the voltage splits unevenly: 36 % on one channel, 64 % on the other, whatever the setpoints. Scaled up, a 48 V pack would put about 31 V on one 30 V channel. So a pack only goes on a pair that is already on, at 0 A with its setpoint above the pack. Measured in that state, the pair passes 60 µA, about as inert as it gets. Unplug the pack before switching the pair off.
  • Channel 3 cannot see that current. When CH3 drives current into CH2, its own readback shows 0 mA: because of the shared negative, the current returns inside the instrument without crossing CH3’s measurement. Until the multimeter went in series, we believed nothing was flowing.

Charging a real pack on the pair is still untested: it passes in simulation, and waits for a pack.

Safety rules

The bench drives real hardware, so a few rules apply to every script and every session, human or not. Know what is wired before turning anything on. Set voltage and current before switching on (at power-up the supply starts at 0 V but with a 3 A limit). Start low. With the load, the source comes on first and goes off last. At the end, outputs off and setpoints back to zero, so that an accidental on delivers nothing.

What next

Expose the supply’s 2048-step timer and the load’s dynamic and list modes in 462lxi; read the multimeter over its infrared link; and add a relay board, driven over the network, to disconnect outputs instead of relying on an inert state that is not quite inert.

Written by Jonathan (462engineering) and Claude (Claude Opus 5.5, Anthropic) — research, debugging and drafting done together in a Claude Code session.

← back to 2026-10-31