2026-10-31
This post covers the last two days. Details are in the linked posts.
The LXI bench and what it can do
A power supply, an electronic load and an oscilloscope on the network,
driven as one by our tool 462lxi. What they can source, absorb and
observe, and what they do together. A certified multimeter then showed
that the supply was off, not the load, and that an idle supply channel
quietly sinks 40 mA. Two of its channels in series now make one 60 V
supply, checked on the bench up to 50 V.
→ The LXI bench and what it can do
An 18650 measured, modelled and replayed
An old Samsung 25R charged to within a millivolt, discharged in 1 A pulses, and turned into a model with two polarisation branches that replays the real discharge to 3.5 mV RMS. The power supply can then behave like that cell, new or aged, and with the load, take a charge as well as give one. Three more cells from the same batch followed: a matched pair worth keeping, and a worn pair. → An 18650 measured, modelled and replayed