Short version: our Rigol DL3031A electronic load had a dead short on its input, whatever its settings. The power MOSFETs were fine: one pair was being driven fully on by its op-amp, an MC33078 that turned out to be dead. The load now runs without that pair until new parts arrive. Meanwhile, our bench tool 462lxi learned to drive it, and to correct its readings in software after a sweep against the power supply.

The symptom

The input measured about 40 mΩ, purely resistive, all the time. Input on or off, any mode, any setpoint, after a reset, after a firmware update: no difference. An electronic load that shorts its source as soon as it is connected is worse than useless, since it can damage what it is meant to test.

The power stage

The current is absorbed by IRFP250N MOSFETs working in their linear region. They are organised in pairs, and each pair has its own MC33078 dual op-amp, one channel per MOSFET, which sets the gate voltage so that the current follows the setpoint. The current is read on a single 0.5 mΩ shunt, followed by a precision amplifier (ADA4522-1).

A short on the input therefore has two possible causes. Either a MOSFET has failed short, or something is holding a gate high and the MOSFET is just doing what it is told.

Following the gates

With the load powered and its input off, every gate should be at zero. We measured the gate-source voltage of each MOSFET: all at zero, except both MOSFETs of one pair, at 8 to 9 V. That is fully on, and two IRFP250N fully on in parallel give roughly the resistance we measured.

Both channels of the same chip at the same wrong voltage pointed to the chip, not to the transistors. We desoldered the MC33078 of that pair, and the short disappeared. With the pair’s gates left without a driver, we ran the usual battery of tests: CC, CV, CR, CP, input on and off. Everything passed, with current readings within ±5 mA of the power supply.

Confirming the culprit

A fault that disappears when a part is removed does not prove that the part is guilty: something around it could have been driving it wrong. So we tested the MC33078 on its own, wired as two voltage followers, powered at ±15 V by two channels of the bench supply. Whatever the input, both outputs stayed stuck at the positive rail. The chip is dead.

Before fitting a new one, we will check that its surroundings did not kill it: continuity of the supply pads (pins 4 and 8) to the ±15 V rails, and the gate resistor and zener of both channels. New MC33078s and a few spare IRFP250Ns are on order. Until the pair is back, the load stays well below its rating.

A load in 462lxi

462lxi is the small command line tool we use to drive the bench instruments over the network (a Rigol DP832 supply and an MSO1000Z oscilloscope), in plain Python with no VISA library. It now drives the load as well:

462lxi load cc 1.5A           # constant current, 1.5 A
462lxi load on
462lxi load meas -w           # U / I / P / R, refreshed
462lxi load meas --raw        # readings without correction
462lxi load sense on          # measure at the Sense terminals

The load has a dangerous habit: any range change resets the setpoint of the mode to zero, and in CR mode a zero resistance is a short. So 462lxi only changes mode or range with the input off, and brings the setpoint to its minimum first. It also reads back the range after setting it, because :SOURce:CURRent:RANGe is silently ignored outside CC mode. Setpoints are capped at the rated values (150 V, 60 A, 350 W), although the instrument accepts slightly more.

Measuring the error

Testing the repair showed that the load’s readings were off. To know by how much, tools/dload_sweep.py connects the load to channel 2 of the DP832, with the load’s Sense terminals wired to the supply’s output, so that both instruments measure the voltage at the same point. The supply is the reference.

For each point, the script waits, takes six readings at 0.5 s intervals on both instruments, and starts again (up to three times) if the first half differs from the second. It stops at once if the load reports a protection or turns its input off by itself: a point measured after that would be wrong and nothing would show it. A few precautions keep the sequence safe: the source is off between blocks (turning the load on in front of a charged output capacitor would empty it in one go); in CV mode the load is turned on first and the source rises with a current limit; and since the load does not start on a setpoint of 5 mA or less, it starts at about 20 mA and is then brought down to the point.

169 points later, in all modes and ranges, the picture was clear:

  • the voltage reads low by about 7 mV + 0.11 %, in every mode;
  • the current reads low by about 2 mA almost everywhere;
  • except in CV mode, 15 V range, where it reads 43.7 mA too high, constant. It is there even with no current at all: input on in CV 1 V, nothing flowing, the load reads +42.7 mA.

We do not know yet whether this offset is older than the failure. The sweep will be run again once the pair is refitted.

Why in software

The obvious fix would be to recalibrate the instrument. We read its calibration data (:CALibration:DATA:GET? and friends), but writing it back is another matter: Rigol’s procedure is not public, the command that opens calibration mode accepts any argument without any way to check what state it is in, and there is no safe way to rewrite the tables. Breaking the internal calibration of a working instrument to fix a 43 mA offset is a bad trade.

So 462lxi corrects the readings itself. The model is linear, read = true × (1 + slope) + offset, with one set of coefficients for the voltage, one for the current in CV 15 V, and one for the current elsewhere. Each set also has a floor: below it, the load reads a clean zero (input off), and the correction leaves it alone instead of turning it into minus the offset. After correction, current and voltage fall within 1 to 2 mA and a few mV of the supply.

The correction is on by default, and --raw gives the instrument’s own values. It assumes that the DP832 is right, which is an assumption and not a traceable calibration, and that the load does not drift: if anything changes, the sweep has to be run again and the coefficients recomputed. The script, the data and the procedure are kept with the tool, so that takes about twenty minutes.

What next

Fit the new MC33078 after checking its surroundings, run the full test battery on all the MOSFETs, then repeat the sweep to see whether the CV 15 V offset survives the repair.

Written by Jonathan (462engineering) and Claude (Claude Opus 5.5, Anthropic) — research, debugging and drafting done together in a Claude Code session.